Features of GE Apex Pro Telemetry:
The ApexPro telemetry receiver features a compact design suitable for mounting to a standard network equipment rack or shelf. Its configuration allows for up to four quad receiver modules, accommodating a total of 16 multi-channel telemetry receivers within a single unit. This flexibility enables the system to expand in tandem with the number of patients undergoing telemetry monitoring.
Multi-channel receivers interface with the Clinical Information Center CIC Pro. Telemetry data undergoes processing, storage, and display at this central location. The ECG data acquired telemetrically can also be viewed on a bedside monitor. Utilizing LAN data transfer technology allows for the receiver system’s versatile placement within the hospital. LED displays facilitate quick online status assessment, while diagnostic ports expedite troubleshooting.
Additional Specifications for GE Apex Pro Telemetry:
**Dimensions**
– Size: H x W x D: 170mm (6.7 in) x 325 mm (12.8 in) x 250 mm (9.8in)
– Weight: 6.4 kg (14 lb)
**Battery**
– Type: ANSI/NEDA 15A, 1.5 V AA alkaline (2 req.)
– Life: 40 hours typical
– Polarity: Electronic reverse polarity protection
**Alarm & Controls**
– Battery integrity: Transmitted and indicated via LED
– Lead fail indication: Transmitted and indicated via LED
– Alarm Pause: Transmitted and indicated via LED
– Graph request: Transmitted
– Power on/off: Battery insertion/removal
**Transmission**
– Frequency stability: ± 0.0001% of assigned channel frequency
– Modulation: GMSK
– Bit rate: 10kb/sec
– Dual antenna: Formed by leadwire shield and internal microstrip antenna
**ECG**
– Multi-channel (5- or 6-leadwire) configuration: I, II, III, Va, Vb, aVR, aVL, aVF
– Leads analyzed simultaneously: Four (I, II, III, and V)
– Single-channel (3-leadwire) configuration: I, II, or III, programmable
– Heart rate detection range: 3 to 300 beats/minute
– QRS detection range: 0.5 to 5 mV
– Frequency response: 0.05Hz to 40Hz (-3dB)
– Dynamic range: ± 5 mV (RTI)
– Input offset: ± 400 mV (RTI)
– Input impedance: 15 Megaohm min differential at 10 Hz
**Analog/Digital**
– A/D converter resolution: 10 bits, 9.76 ?V (RTI)
– Sample rate: 240 samples/sec
– Serial communications: 2-9600 baud asynchronous
**Operating Conditions**
– Ambient temperature: 0°C to 50°C (32°F to 122°F)
– Relative humidity: 5% to 95% (noncondensing)
**Storage Conditions**
– Temperature: -40°C to 70°C (-40°F to 158°F)
– Relative humidity: 30% to 95% (noncondensing)
– Pressure: 700 hPa to 1060 hPa
**Quad Receiver Module**
– Type: GMSK digitally demodulated
– UHF frequency range: 560.025 MHz to 613.975 MHz (U.S.) 420 MHz to 460 MHz (International)
– Frequency step resolution: Frequency synthesized tuning to any transmitter. 25 kHz spacing
– Frequency stability: ± 0.00015% of assigned channel frequency
– Demodulation: GMSK
– Bit rate: 10 kb/sec
– Sensitivity: 4?V (-95dBm) minimum for 1 bit error/1 million bits received
**Receiver System**
– Capacity: 1 to 4 quad receiver modules (4 to 16 receivers)
– System status indicators: 7 bicolor LEDs
**Network**
– IEEE 802.3 compatible, physical connector via 10 Base T
– Serial diagnostics: 19200 baud, 1 stop bit, 8 data bits, no parity, XON/XOFF flow control
**Power Requirements**
– Line Voltage: 85 to 264 VAC, 47 to 63 Hz
– Power consumption: 25 watts max with 4 quad receiver units
– Cooling: Free convection
Reviews
There are no reviews yet.